Recommended Essential Steps in a Pre-Anesthesia Checkout Procedure

TO BE COMPLETED DAILY, OR AFTER A MACHINE IS MOVED OR VAPORIZERS

CHANGED ITEM TO BE COMPLETED RESPONSIBLE PARTY

Item #1:       Verify Auxiliary Oxygen Cylinder and Manual Ventilation Device (Ambu Bag) are Available & Functioning.

Item #2:       Verify patient suction is adequate to clear the airway.

Item #3:       Turn on anesthesia delivery system and confirm that ac power is available

Item #4:       Verify availability of required monitors, including alarms

Item #5:       Verify that pressure is adequate on the spare oxygen cylinder mounted on the anesthesia machine

Item #6:       Verify that the piped gas pressures are ³ 50 psig.

Item #7:       Verify that vaporizers are adequately filled and, if applicable, that the filler ports are tightly closed.        Provider or Tech

Item #8:       Verify that there are no leaks in the gas supply lines between the flowmeters and the common gas outlet.

Item #9:       Test scavenging system function.

Item #10:     Calibrate, or verify calibration of, the oxygen monitor and check the low oxygen alarm

Item #11:     Verify carbon dioxide absorbent is fresh and not exhausted.

Item #12:     Perform breathing system pressure and leak testing.      

Item #13:     Verify that gas flows properly through the breathing circuit during both inspiration and exhalation.      

Item #14:     Document completion of checkout procedures

Item #15:     Confirm ventilator settings and evaluate readiness to deliver anesthesia care. (ANESTHESIA TIME OUT)

 

http://www.apsf.org/assets/Documents/summer2008.pdf